Can Computer Tomography Predict Compromise of Cardiac Structures After Percutaneous Closure of Interatrial Septal Defects?
Abstract
Background: Erosion of a cardiac structure after device closure of an interatrial septal communication (IASC-C), although rare, is a major and severe adverse event which may be underreported. On the other hand, unexplained episodes of transient chest pain occur more often and may be quite distressing. We sought to define the parameters relating the devices to the adjacent cardiac structures and to determine whether computer tomography (CT) could predict erosion of atrial or aortic wall or precordial pain symptoms occurring in the first months after device implantation.
Methods: Retrospective observational study of 20 patients who underwent CT for de novo chest pain occurring after IASC-C or as a diagnostic test for suspected or proven coronary artery disease (CAD). Clinical follow up was for 20.5 17.6 (6-84) months. CT was done 18 10 (2-28) weeks after IASC-C.
Results: Indentation of the aortic root was found in 11 (55%) patients, the left atrial wall in 13 (65%) and the right atrial wall in eight (40%) of patients. Contact without indentation was found in nine (45%), 6 (30%) and 11 (55%) of patients respectively.
Conclusions: Device indenting of the left and right atrial, as well as the aortic wall, occured in the majority of the patients examined after IASC-C. This finding may explain bouts of chest pain after the intervention in some patients, but does not predict clinically relevant erosion of a cardiac structure.
Cardiol Res. 2011;2(6):288-292
doi: https://doi.org/10.4021/cr109w
Methods: Retrospective observational study of 20 patients who underwent CT for de novo chest pain occurring after IASC-C or as a diagnostic test for suspected or proven coronary artery disease (CAD). Clinical follow up was for 20.5 17.6 (6-84) months. CT was done 18 10 (2-28) weeks after IASC-C.
Results: Indentation of the aortic root was found in 11 (55%) patients, the left atrial wall in 13 (65%) and the right atrial wall in eight (40%) of patients. Contact without indentation was found in nine (45%), 6 (30%) and 11 (55%) of patients respectively.
Conclusions: Device indenting of the left and right atrial, as well as the aortic wall, occured in the majority of the patients examined after IASC-C. This finding may explain bouts of chest pain after the intervention in some patients, but does not predict clinically relevant erosion of a cardiac structure.
Cardiol Res. 2011;2(6):288-292
doi: https://doi.org/10.4021/cr109w
Keywords
ASD; PFO; Percutaneous closure; Complications; Computer tomography