Cardiol Res
Cardiology Research, ISSN 1923-2829 print, 1923-2837 online, Open Access
Article copyright, the authors; Journal compilation copyright, Cardiol Res and Elmer Press Inc
Journal website http://www.cardiologyres.org

Case Report

Volume 6, Number 2, April 2015, pages 257-259


Wondering PTCA Wire: Retrieval by Tangling Technique

Santosh Kumar Sinhaa, b, Chandra Mohan Vermaa, Vinay Krishnaa, Ramesh Thakura, Prakash Kumara, Amit Goela, Ashutosh Kumara, Mahmadulla Razia

aDepartment of Cardiology, LPS Institute of Cardiology, G.S.V.M. Medical College, G. T. Road, Kanpur, Uttar Pradesh 208002, India
bCorresponding Author: Santosh Kumar Sinha, Department of Cardiology, LPS Institute of Cardiology, G.S.V.M. Medical College, G. T. Road, Kanpur, Uttar Pradesh 208002, India

Manuscript accepted for publication April 15, 2015
Short title: Wondering PTCA Wire
doi: http://dx.doi.org/10.14740/cr392w

Abstract▴Top 

A 38-year-old man underwent coronary angiography in our institution due to acute myocardial infarction as part of pharmaco-invasive strategy following thrombolysis. The patient showed total occlusion of mid left anterior descending (LAD) artery which was tortuous and calcified. The planned treatment was percutaneous coronary intervention (PCI) of culprit artery with wire being “jailed” in obtuse marginal branch of left circumflex artery (LCX) as left main was short and because of lesion characteristics. After successful stent implantation in the LAD, the “jailed” wire fractured as guiding catheter got deeply intubated as stent was being deployed in LAD. Initially, two balance middle weight (BMW) wires were used to retrieve but failed. Wire was wondering as it moved to proximal LCX, left main, partly into aortic sinus, sometimes proximal LAD and finally to LCX again during retrieval. Then it was decided to use tangling technique with the help of three BMW wires acting as rescue wires. The proximal ends of all three wires were inserted together in a torque device which were firmly screwed and rotated 40 - 50 times in circular pattern. During this rotational motion, the broken segment was tangled within these rescue wires and all four wires were removed together.

Keywords: Jailed wire; Rescue wire; Tangling technique; Wondering wire

Introduction▴Top 

Virtually every coronary angioplasty device is advanced into the coronary system over a wire. The soft, atraumatic tips of coronary wires have been known to fracture off, if being manipulated excessively, and embolize in the coronary circulation. This most frequently occurs when the shapeable wire tip becomes lodged in an atheromatous plaque and separates from the body of the wire when the wire is retracted or entrapment, overcoiling and excessive traction of the guide wire can cause it to fracture. Because of manufacturing design, the incidence was a little high in the past but as current coronary wires are constructed of gradually tapering filaments, so relatively weak junction are minimized in contemporary wire design. Nevertheless it still occurs with incidence around 0.2-0.8% [1]. If the fragment of guide wire is not removed, there is a likelihood of acute embolization leading to acute coronary thrombus. Our case was caused by a deep intubation of guiding catheter into LAD as vessel was tortuous and long stent was being delivered leading to fracture of jailed wire of LCX. This type of complication is rarely reported as jailed wire kept wondering from proximal LCX, left main, partly into aortic sinus, sometimes proximal LAD and finally to LCX again during retrieval which was then successfully retrieved by tangling technique with the help of three BMW wires acting as rescue wires.

Case Report▴Top 

A 38-year-old man with diabetes and hypertension of 2-year duration was admitted with acute anterior wall myocardial infarction for which he was thrombolyzed with 30 mg tenecteplase. He underwent coronary artery angiography on following day. PCI was planned because of total occlusion of mid LAD artery (Fig. 1). Intravenous heparin (100 U/kg) was administered. Left main was hooked with 6 F Launcher® Coronary Guide Catheter (Medtronic, USA). Lesion was crossed by Fielder wire (Asahi Intecc, Japan) which was passed through FineCrossTM MG microcatheter (Terumo Medical Corp., NJ, USA) and parked distally. Wire was exchanged with 0.014ʺ 190 cm Hi-Torque BMW (Abbott Vascular, Santa Clara, USA) and microcatheter was removed by Nanto’s technique. Lesion was repeatedly dilated with 1.25 × 6, 1.5 × 10, 2 × 10 and 2.5 × 10 Sprinter Legend RX Balloon (Medtronic, USA). As artery appeared tortuous and lesion was calcified, it was planned to stent LAD with jailing the BMW wire into obtuse marginal branch of LCX to have better support and tracking of stent. Xience Prime 2.75 × 30 mm (Abbott Vascular, Santa Clara, USA) was deployed in distal LAD at 12 atm. Xience Prime 3 × 28 mm was deployed in proximal LAD overlapping with distal stent at 13 atm with post procedural TIMI III flow (Fig. 2). During proximal stent deployment, guiding catheter got deeply intubated causing fracture of “jailed” wire (Fig. 3). Two BMW wires were parked in distal LCX and proximal ends of both wires were passed through torque device and rotated 40 - 50 times to catch hold it wire (Fig. 4). While pulling it slipped in left main, one end was floating in aortic sinus and other was in proximal LCX (Fig. 5). Similar maneuver was applied but this time it migrated to proximal LAD (Fig. 6) and finally to distal LCX again. Then it was decided to use three BMW wires which were parked in distal LCX and proximal ends of all three wires were passed through torque device and rotated 40 - 50 times to catch hold it wire and all four wires were pulled inside guiding catheter and successfully retrieved (Fig. 7).

Figure 1.
Click for large image
Figure 1. Antero-posterior (AP) cranial view showing total occlusion of mid left anterior descending (LAD) artery.

Figure 2.
Click for large image
Figure 2. Antero-posterior (AP) cranial view showing post procedural TIMI III flow of left anterior descending (LAD) artery.

Figure 3.
Click for large image
Figure 3. Antero-posterior (AP) cranial view showing broken distal segment of jailed wire in obtuse marginal branch of left circumflex artery (LCX).

Figure 4.
Click for large image
Figure 4. Antero-posterior (AP) caudal view showing distal segment of jailed wire being entangled by two BMW wires.

Figure 5.
Click for large image
Figure 5. Antero-posterior (AP) caudal view showing one free end in aortic sinus and other end in proximal left circumflex artery (LCX).

Figure 6.
Click for large image
Figure 6. Antero-posterior (AP) caudal view showing broken wire in proximal left anterior descending (LAD) artery and other end in proximal LCX.

Figure 7.
Click for large image
Figure 7. Antero-posterior (AP) caudal view showing distal segment of jailed wire being entangled by triple rescue wires.
Discussion▴Top 

It is generally believed that the retained guide wire in the coronary tree may result in complications such as emboli, thrombosis, dissection and rupture and should be removed from the coronary system [2, 3]. Various solutions have been described to solve the problem of a broken guide wire in a coronary artery, ranging from surgery to simply leaving the segment of wire in place in cases where it could not be removed but first, efforts should be made to remove the guide wire percutaneously [4, 5]. However, it can be left as such in vessels that are already occluded or in very small branches. In our case as obtuse margin was quite large and LCX was dominant, we decided to retrieve it in any way. As snare loop wire was not available, simple technique was tried. Two wires were used first but it slipped in proximal LCX and then it migrated to left main, proximal LAD, once floating into aortic sinus and finally to obtuse marginal branch again. Finally tangling technique with the help of three wires acting as rescue wires was used to successfully retrieve the broken wire. Usually wire breaks in the artery which is being tackled for various reasons already described but our case was unique as jailed wire of LCX got broken by the guiding catheter at the time when proximal stent was being deployed in LAD. To the best of our knowledge, it is being reported for the first time.


References▴Top 
  1. Chang TM, Pellegrini D, Ostrovsky A, Marrangoni AG. Surgical management of entrapped percutaneous transluminal coronary angioplasty hardware. Tex Heart Inst J. 2002;29(4):329-332.
    pubmed
  2. Hartzler GO, Rutherford BD, McConahay DR. Retained percutaneous transluminal coronary angioplasty equipment components and their management. Am J Cardiol. 1987;60(16):1260-1264.
    doi
  3. Sethi GK, Ferguson TB, Jr., Miller G, Scott SM. Entrapment of broken guidewire in the left main coronary artery during percutaneous transluminal coronary angioplasty. Ann Thorac Surg. 1989;47(3):455-457.
    doi
  4. Patel T, Shah S, Pandya R, Sanghvi K, Fonseca K. Broken guidewire fragment: a simplified retrieval technique. Catheter Cardiovasc Interv. 2000;51(4):483-486.
    doi
  5. Demircan S, Yazici M, Durna K, Yasar E. Intracoronary guidewire emboli: a unique complication and retrieval of the wire. Cardiovasc Revasc Med. 2008;9(4):278-280.
    doi pubmed


This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Cardiology Research is published by Elmer Press Inc.

 

Browse  Journals  

 

Journal of Clinical Medicine Research

Journal of Endocrinology and Metabolism

Journal of Clinical Gynecology and Obstetrics

 

World Journal of Oncology

Gastroenterology Research

Journal of Hematology

 

Journal of Medical Cases

Journal of Current Surgery

Clinical Infection and Immunity

 

Cardiology Research

World Journal of Nephrology and Urology

Cellular and Molecular Medicine Research

 

Journal of Neurology Research

International Journal of Clinical Pediatrics

 

 
       
 

Cardiology Research, bimonthly, ISSN 1923-2829 (print), 1923-2837 (online), published by Elmer Press Inc.                     
The content of this site is intended for health care professionals.

This is an open-access journal distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License, which permits unrestricted
non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Creative Commons Attribution license (Attribution-NonCommercial 4.0 International CC-BY-NC 4.0)


This journal follows the International Committee of Medical Journal Editors (ICMJE) recommendations for manuscripts submitted to biomedical journals,
the Committee on Publication Ethics (COPE) guidelines, and the Principles of Transparency and Best Practice in Scholarly Publishing.

website: www.cardiologyres.org   editorial contact: editor@cardiologyres.org
Address: 9225 Leslie Street, Suite 201, Richmond Hill, Ontario, L4B 3H6, Canada

© Elmer Press Inc. All Rights Reserved.


Disclaimer: The views and opinions expressed in the published articles are those of the authors and do not necessarily reflect the views or opinions of the editors and Elmer Press Inc. This website is provided for medical research and informational purposes only and does not constitute any medical advice or professional services. The information provided in this journal should not be used for diagnosis and treatment, those seeking medical advice should always consult with a licensed physician.